Featuring a real-time clock, this handheld thermometer readout stores up to 15,000 time- and date-stamped measurements.
Featuring a real-time clock, this handheld thermometer readout stores up to 15,000 time- and date-stamped measurements.
Discontinued!
This product has been discontinued and is no longer available.
This is the direct replacement:
This thermometer readout delivers exceptional accuracy, wide measurement range, logging and trending, all in a handheld tool you can take anywhere.
Features
Input Channels | 2 |
Logging | 15,000 time and date stamped; 25 readings with statistics |
Sample Interval (Normal) | 1 second |
Sample Interval (Fast Mode) | 0.3 seconds (see technical manual for details) |
Sensor Types | PRTs, RTDs, thermistors, thermocouples |
Thermocouple Types | C, E, J, K, L, M, N, T, U, B, R, S |
Operating Temperature | 14 to 140°F (-10 to 60°C) (best accuracy 55.4 to 91.4°F (13 to 33°C)) |
Power Requirements | 3 AA alkaline batteries |
Size | 3.75 x 7.9 x 1.86in (96 x 200 x 47mm) |
Environmental Conditions For Best Accuracy | 55.4 to 91.4°F (13 to 33°C) |
Millivolt Range And Accuracy | -10 to 75mV ±(0.005% + 5µV) |
Resistance Range And Accuracy | 0 to 400Ω ±(0.004% + 0.002Ω); 200 to 50kΩ ±(0.01% + 0.5Ω); 50 to 500kΩ ±(0.03%) |
Temperature Coefficient, Voltage (-10 to 13°C, +33 to 60°C) | ±(0.001%/°C + 1µV/°C) |
Temperature Coefficient, Resistance (-10 to 13°C, +33 to 60°C) | 0.0008%/°C + 0.0004Ω (0 to 400Ω); 0.002%/°C + 0.1Ω (0 to 50kΩ); 0.06%/°C + 0.1Ω (50 to 500kΩ) |
Excitation Current, Resistance | 1mA (0 to 400Ω); 10µA (0 to 50kΩ); 2µA (50 to 500kΩ) |
Click on a category to view a selection of compatible accessories with the Fluke 1524-P1-156 Handheld Thermometer Readout with 5627A PRT.
Input Channels | 2 |
Logging | 15,000 time and date stamped; 25 readings with statistics |
Sample Interval (Normal) | 1 second |
Sample Interval (Fast Mode) | 0.3 seconds (see technical manual for details) |
Sensor Types | PRTs, RTDs, thermistors, thermocouples |
Thermocouple Types | C, E, J, K, L, M, N, T, U, B, R, S |
Operating Temperature | 14 to 140°F (-10 to 60°C) (best accuracy 55.4 to 91.4°F (13 to 33°C)) |
Power Requirements | 3 AA alkaline batteries |
Size | 3.75 x 7.9 x 1.86in (96 x 200 x 47mm) |
Environmental Conditions For Best Accuracy | 55.4 to 91.4°F (13 to 33°C) |
Millivolt Range And Accuracy | -10 to 75mV ±(0.005% + 5µV) |
Resistance Range And Accuracy | 0 to 400Ω ±(0.004% + 0.002Ω); 200 to 50kΩ ±(0.01% + 0.5Ω); 50 to 500kΩ ±(0.03%) |
Temperature Coefficient, Voltage (-10 to 13°C, +33 to 60°C) | ±(0.001%/°C + 1µV/°C) |
Temperature Coefficient, Resistance (-10 to 13°C, +33 to 60°C) | 0.0008%/°C + 0.0004Ω (0 to 400Ω); 0.002%/°C + 0.1Ω (0 to 50kΩ); 0.06%/°C + 0.1Ω (50 to 500kΩ) |
Excitation Current, Resistance | 1mA (0 to 400Ω); 10µA (0 to 50kΩ); 2µA (50 to 500kΩ) |
Click on a category to view a selection of compatible accessories with the Fluke 1524-P1-156 Handheld Thermometer Readout with 5627A PRT.