Equipped with 4-wire connections, this secondary PRT makes less noisy measurements when compared to its 2-wire counterparts.
Equipped with 4-wire connections, this secondary PRT makes less noisy measurements when compared to its 2-wire counterparts.
Features
Temperature Range | -328 to 932°F (-200 to 500°C) |
Nominal Resistance at 0.01°C | 100Ω ±0.5Ω |
Temperature Coefficient | 0.0039250Ω/Ω/°C |
Short-Term Repeatability | ±0.01 at 0.010°C; ±0.02°C at max temp |
Drift | ±0.01 at 0.010°C; ±0.02°C at max temp |
Hysteresis | ±0.01°C maximum |
Sensor Length | 1.2in ±0.2in (30mm ±5mm) |
Sensor Location | 0.1in ± 0.1in (3mm ±1mm from tip) |
Minimum Insulation Resistance | 500MΩ at 23°C, 20MΩ at 500°C |
Transition Junction Dimensions | 2.8 x 0.49in (71 x 12.5mm) |
Minimum Immersion Length (<5 mK Error) | 3.1in (80mm) |
Maximum Immersion Length | 12in (305mm) |
Response Time | 9 seconds typical |
Self Heating (In 0°C Bath) | 75mW/°C |
Lead-Wire Cable Type | Teflon, 24AWG |
Lead-Wire Length | 6ft (1.8m) |
Lead-Wire Temperature Range | -58 to 482°F (-50 to 250°C) |
Calibration | Calibration not included; NVLAP-accredited calibration optional, lab code 200348-0; please see calibration uncertainty table and its explanation of changeable uncertainties |
Click on a category to view a selection of compatible accessories with the Fluke 5608-9-M Secondary PRT with Mini Banana Plugs, -200 to 500°C, Calibration Options.
Temperature Range | -328 to 932°F (-200 to 500°C) |
Nominal Resistance at 0.01°C | 100Ω ±0.5Ω |
Temperature Coefficient | 0.0039250Ω/Ω/°C |
Short-Term Repeatability | ±0.01 at 0.010°C; ±0.02°C at max temp |
Drift | ±0.01 at 0.010°C; ±0.02°C at max temp |
Hysteresis | ±0.01°C maximum |
Sensor Length | 1.2in ±0.2in (30mm ±5mm) |
Sensor Location | 0.1in ± 0.1in (3mm ±1mm from tip) |
Minimum Insulation Resistance | 500MΩ at 23°C, 20MΩ at 500°C |
Transition Junction Dimensions | 2.8 x 0.49in (71 x 12.5mm) |
Minimum Immersion Length (<5 mK Error) | 3.1in (80mm) |
Maximum Immersion Length | 12in (305mm) |
Response Time | 9 seconds typical |
Self Heating (In 0°C Bath) | 75mW/°C |
Lead-Wire Cable Type | Teflon, 24AWG |
Lead-Wire Length | 6ft (1.8m) |
Lead-Wire Temperature Range | -58 to 482°F (-50 to 250°C) |
Calibration | Calibration not included; NVLAP-accredited calibration optional, lab code 200348-0; please see calibration uncertainty table and its explanation of changeable uncertainties |