Designed to perform in extreme environments, this secondary PRT comes equipped with mini banana plug connections.
Designed to perform in extreme environments, this secondary PRT comes equipped with mini banana plug connections.
Features
Temperature Range | -328 to 1,238°F (-200 to 670°C) |
Nominal Resistance at 0.01°C | 100Ω ±0.5Ω |
Temperature Coefficient | 0.0039250Ω/Ω/°C |
Short-Term Repeatability | ±0.01 at 0.010°C; ±0.02°C at max temp |
Drift | ±0.01 at 0.010°C; ±0.02°C at max temp |
Hysteresis | ±0.01°C maximum |
Sensor Length | 1.2in ±0.2in (30mm ±5mm) |
Sensor Location | 0.1in ±0.1in (3mm ±1mm from tip) |
Minimum Insulation Resistance | 500MΩ at 23°C, 10MΩ at 670°C |
Transition Junction Dimensions | 2.8 x 0.49in (71 x 12.5mm) |
Minimum Immersion Length (<5 mK Error) | 3.9in (100mm) |
Maximum Immersion Length | 12in (305mm) |
Response Time | 12 seconds typical |
Self Heating (In 0°C Bath) | 50mW/°C |
Lead-Wire Cable Type | Teflon, 24AWG |
Lead-Wire Length | 6ft (1.8m) |
Lead-Wire Temperature Range | -58 to 482°F (-50 to 250°C) |
Calibration | Calibration not included; NVLAP-accredited calibration optional, lab code 200348-0; please see calibration uncertainty table and its explanation of changeable uncertainties |
Click on a category to view a selection of compatible accessories with the Fluke 5609-12-M Secondary PRT with Mini Banana Plugs, -200 to 670°C, Calibration Options.
Temperature Range | -328 to 1,238°F (-200 to 670°C) |
Nominal Resistance at 0.01°C | 100Ω ±0.5Ω |
Temperature Coefficient | 0.0039250Ω/Ω/°C |
Short-Term Repeatability | ±0.01 at 0.010°C; ±0.02°C at max temp |
Drift | ±0.01 at 0.010°C; ±0.02°C at max temp |
Hysteresis | ±0.01°C maximum |
Sensor Length | 1.2in ±0.2in (30mm ±5mm) |
Sensor Location | 0.1in ±0.1in (3mm ±1mm from tip) |
Minimum Insulation Resistance | 500MΩ at 23°C, 10MΩ at 670°C |
Transition Junction Dimensions | 2.8 x 0.49in (71 x 12.5mm) |
Minimum Immersion Length (<5 mK Error) | 3.9in (100mm) |
Maximum Immersion Length | 12in (305mm) |
Response Time | 12 seconds typical |
Self Heating (In 0°C Bath) | 50mW/°C |
Lead-Wire Cable Type | Teflon, 24AWG |
Lead-Wire Length | 6ft (1.8m) |
Lead-Wire Temperature Range | -58 to 482°F (-50 to 250°C) |
Calibration | Calibration not included; NVLAP-accredited calibration optional, lab code 200348-0; please see calibration uncertainty table and its explanation of changeable uncertainties |