| Measurement Accuracy |
| Pressure |
| Fluke 718Ex 30G | | | Range | -830 mbar to 2 bar (internal sensor) | | Over pressure: 3x FS | | Resolution | 0.0001 bar | | Accuracy | 0.05% FS | | Functions | Zero, Min, Max, Hold, Damp | | Note: for use with non-corrosive gasses | | Fluke 718Ex 100G | | | Range | -830 mbar to 7 bar (internal sensor) | | Over pressure: 2x FS | | Resolution | 0.001 bar | | Accuracy | 0.05% FS | | Functions | Zero, Min, Max, Hold, Damp | | Note: for use with non-corrosive gasses | | Fluke 718Ex series | | | via pressure module connector | | Range | 29 pressure modules, 2.5 mbar to 700 bar | | Over pressure per pressure module specs | | Resolution | Per pressure module specs | | Accuracy | To 0.025% of full span, per pressure module specs | | Functions | Zero, Min, Max, Hold, Damp | | Note: media compatibility per pressure module specs | | Source Pressure, built in pump | | | Range | -820 mbar to full scale | | Note | Supported Pressure Units PSI, in. H2O (4°C), in. H2O (20°C), kPa, cm H2O (4°C), | | cm H2O (20°C), bar, mbar, kg/cm2, mmHg, in Hg | |
| Fluke 718Ex 30G |
|
| Range |
-830 mbar to 2 bar (internal sensor) |
| |
Over pressure: 3x FS |
| Resolution |
0.0001 bar |
| Accuracy |
0.05% FS |
| Functions |
Zero, Min, Max, Hold, Damp |
| |
Note: for use with non-corrosive gasses |
| Fluke 718Ex 100G |
|
| Range |
-830 mbar to 7 bar (internal sensor) |
| |
Over pressure: 2x FS |
| Resolution |
0.001 bar |
| Accuracy |
0.05% FS |
| Functions |
Zero, Min, Max, Hold, Damp |
| |
Note: for use with non-corrosive gasses |
| Fluke 718Ex series |
|
| |
via pressure module connector |
| Range |
29 pressure modules, 2.5 mbar to 700 bar |
| |
Over pressure per pressure module specs |
| Resolution |
Per pressure module specs |
| Accuracy |
To 0.025% of full span, per pressure module specs |
| Functions |
Zero, Min, Max, Hold, Damp |
| |
Note: media compatibility per pressure module specs |
| Source Pressure, built in pump |
|
| Range |
-820 mbar to full scale |
| Note |
Supported Pressure Units PSI, in. H2O (4°C), in. H2O (20°C), kPa, cm H2O (4°C), |
| |
cm H2O (20°C), bar, mbar, kg/cm2, mmHg, in Hg |
| Technical Data |
| Measure Accuracy |
| Range | 0-24 mA | | Resolution | 0.001 mA | | Accuracy | 0.02% rdg ± 2 counts | |
| Range |
0-24 mA |
| Resolution |
0.001 mA |
| Accuracy |
0.02% rdg ± 2 counts |
| Display |
| LCD, 5 digit pressure and current simultaneous |
|
| |
LCD, 5 digit pressure and current simultaneous |
| Environmental Specifications |
| Operating temperature |
|
| |
-10°C to 55°C |
| Storage temperature |
|
| |
-40°C to 60°C |
| Humidity (Without Condensation) |
| 95% (10°C to 30°C) | | 75% (30°C to 40°C) | | 45% (40°C to 50°C) | | 35% (50°C to 55°C) | |
| |
95% (10°C to 30°C) |
| |
75% (30°C to 40°C) |
| |
45% (40°C to 50°C) |
| |
35% (50°C to 55°C) |
| Operating Altitude |
|
| |
3,000 m max |
| Safety Specifications |
| Safety |
| ATEX II 1G EEx ia IIC T4 compliant | | CSA Certification I.S. Class I, | | Div 1 Groups A-D , T4 | |
| |
ATEX II 1G EEx ia IIC T4 compliant |
| |
CSA Certification I.S. Class I, |
| |
Div 1 Groups A-D , T4 |
| Mechanical & General Specifications |
| Size |
| 216 mm L x 94 mm W x 66 mm H (with holster) |
|
| |
216 mm L x 94 mm W x 66 mm H (with holster) |
| Weight |
| Approx. 992 g (with holster) |
|
| |
Approx. 992 g (with holster) |
| Warranty |
|
| |
1 year |
| Shock |
|
| |
1 meter drop test |
| Vibration |
|
| |
Random, 2 G, 5-500 Hz |
| Battery type |
| One 9-volt alkaline battery ANSI/NEDA 1604A or IEC 6LR619V | | Battery life: 4 to 20 hours, depending on functions used | |
| |
One 9-volt alkaline battery ANSI/NEDA 1604A or IEC 6LR619V |
| |
Battery life: 4 to 20 hours, depending on functions used |