| Measurement Accuracy |
| Pressure |
| 717 Series |
| | Measure Pressure | (internal sensor) | | Accuracy | 0.05% full scale | | Functions | Zero, Min, Max, Hold, Damp |
| Note Gasses / liquids (non-corrosive) | | |
| | Fluke 717 1000G |
| | Range | 0 to 1000 PSI, (0 mbar to 69 bar, 0 to 6894.76 kPa ) | | Resolution | 0.1 PSI, 1 mbar, 0.1 kPa | | Over Pressure | 2 X full scale | | |
| | Fluke 717 Series with Pressure Modules |
| | Measure Pressure | (via pressure module connector using optional Pressure Modules) | | Range | 29 pressure modules, 0-1 inH20 to 10,000 PSI, 2.5 mbar to 700 bar ranges. Over pressure per pressure module specs | | Resolution | Per pressure module specs | | Accuracy | per pressure module specs | | Functions | Zero, Min, Max, Hold, Damp | |
| 717 Series |
|
| Measure Pressure |
(internal sensor) |
| Accuracy |
0.05% full scale |
| Functions |
Zero, Min, Max, Hold, Damp |
| |
Note Gasses / liquids (non-corrosive) |
| |
|
| Fluke 717 1000G |
|
| Range |
0 to 1000 PSI, (0 mbar to 69 bar, 0 to 6894.76 kPa ) |
| Resolution |
0.1 PSI, 1 mbar, 0.1 kPa |
| Over Pressure |
2 X full scale |
| |
|
| Fluke 717 Series with Pressure Modules |
|
| Measure Pressure |
(via pressure module connector using optional Pressure Modules) |
| Range |
29 pressure modules, 0-1 inH20 to 10,000 PSI, 2.5 mbar to 700 bar ranges. Over pressure per pressure module specs |
| Resolution |
Per pressure module specs |
| Accuracy |
per pressure module specs |
| Functions |
Zero, Min, Max, Hold, Damp |
| Current |
| Range: | 0 to 24 mA (-25% to 125%) | | Resolution | 0.001 mA | | Accuracy | 0.015% + 2 counts | |
| Range: |
0 to 24 mA (-25% to 125%) |
| Resolution |
0.001 mA |
| Accuracy |
0.015% + 2 counts |
| Specifications |
| Loop Power |
| Range | 24 V dc | | Accuracy | +/- 10% |
| Note: Drive; 20 mA into 1000 ohms for battery >6.8V; 700 W for battery 5.8 to 6.8 V | |
| Range |
24 V dc |
| Accuracy |
+/- 10% |
| |
Note: Drive; 20 mA into 1000 ohms for battery >6.8V; 700 W for battery 5.8 to 6.8 V |
| Display |
|
| |
Clear 5 digit LCD |
| Environmental Specifications |
| Operating Temperature |
|
| |
-10 �C to 55 ºC |
| Storage Temperature |
|
| |
-40 ºC to 60 ºC |
| Humidity (Without Condensation) |
| 95% (10 ºC to 30 ºC) |
| 75% (30 ºC to 40 ºC) |
| 45% (40 ºC to 50 ºC) |
| 35% (50 ºC to 55 ºC) | |
| |
95% (10 ºC to 30 ºC) |
| |
75% (30 ºC to 40 ºC) |
| |
45% (40 ºC to 50 ºC) |
| |
35% (50 ºC to 55 ºC) |
| Operating Altitude |
|
| |
3,000 M max |
| Safety Specifications |
| Safety |
| CSA C22.2 No. 1010.1:1992 |
|
| |
CSA C22.2 No. 1010.1:1992 |
| Mechanical & General Specifications |
| Size |
| 201 mm L x 98 mm W x 52 mm (7.93in. L x 3.86in. W X 2.06in.) |
|
| |
201 mm L x 98 mm W x 52 mm (7.93in. L x 3.86in. W X 2.06in.) |
| Weight |
| 22 oz, approximately 600G with holster |
|
| |
22 oz, approximately 600G with holster |
| Warranty |
|
| |
3 years |
| Shock & Vibration |
| Random, 2G, 5-500 Hz |
| 1m drop test | |
| |
Random, 2G, 5-500 Hz |
| |
1m drop test |
| Battery type |
| Single 9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline |
|
| |
Single 9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline |